Technical Specifications
Physical Dimensions
Specifications |
Values |
Width |
1.0 in [25.4 mm] |
Length |
2.75 in [69.9 mm] |
Height |
0.72 in [18.3 mm] |
Weight |
1.25 oz. [35.4 g] |
Measurement
Specifications |
Values |
Technique |
Oscillometric |
NIBP Accuracy |
Meets ANSI/AAMI SP10, EN 1060-4, ISO 81060-2 |
Patient Applications |
Adult, Pediatric, Neonatal |
Systolic Range |
- Adult: 30-255 mmHg
- Neonatal: 30-135 mmHg
|
Diastolic Range |
- Adult: 15-220 mmHg
- Neonatal: 15-110 mmHg
|
Pulse Rate Range |
- Adult: 30-240 BPM (Beats per min)
- Neonatal: 40-240 BPM
|
Pulse Rate Accuracy |
±2% or ±3 BPM, whichever is greater |
Transducer Accuracy |
± 3 mmHg over full range in operating conditions |
Operating Modes |
Manual |
Operating Conditions |
0˚C to 55˚C, 15% to 90%, non-condensing humidity |
Storage Conditions |
-20˚C to +65˚C, up to 90% RH non-condensing |
Electrical
Specifications |
Values |
Operating Voltage |
5 VDC ± 5%, 6 VDC ± 5% |
Power Consumption |
Using 5 VDC Nominal
- Sleep: 50 mW
- Idle: 150 mW
- Measurement: 1.0 W
- Inflation: 3.6 W (High)
|
Communication Protocols |
TTL (5V) |
Safety & Regulatory Standards |
IEC 60601-1, IEC/EN80601-2-30, EN60601-1-2:2007, ISO 81060-2:2009 |
Advantage Technologies
Specifications |
Values |
RMT-MAX (Maximum Performance R-wave Motion Tolerance) |
Yes |
TMT-MAX (Maximum Performance Transport Motion Tolerance) |
Yes |
VET-S (Standard Veterinary Monitoring) |
Yes |
Customer Support
Specifications |
Values |
Standard Support |
Yes |
Custom Support |
10 hours* |
Custom Design |
Additional Charge |
Warranty
Specifications |
Values |
Time |
1 year |
Cycles |
30000 |
*ADDITIONAL SUPPORT CHARGED AT STANDARD RATES