Technical Specifications

Physical Dimensions

Specifications Values
Width 1.0 in [25.4 mm]
Length 2.75 in [69.9 mm]
Height 0.72 in [18.3 mm]
Weight 1.25 oz. [35.4 g]

Measurement

Specifications Values
Technique Oscillometric
NIBP Accuracy Meets ANSI/AAMI SP10, EN 1060-4, ISO 81060-2
Patient Applications Adult, Pediatric, Neonatal
Systolic Range
  • Adult: 30-255 mmHg
  • Neonatal: 30-135 mmHg
Diastolic Range
  • Adult: 15-220 mmHg
  • Neonatal: 15-110 mmHg
Pulse Rate Range
  • Adult: 30-240 BPM (Beats per min)
  • Neonatal: 40-240 BPM
Pulse Rate Accuracy ±2% or ±3 BPM, whichever is greater
Transducer Accuracy ± 3 mmHg over full range in operating conditions
Operating Modes Manual
Operating Conditions 0˚C to 55˚C, 15% to 90%, non-condensing humidity
Storage Conditions -20˚C to +65˚C, up to 90% RH non-condensing

Electrical

Specifications Values
Operating Voltage 5 VDC ± 5%, 6 VDC ± 5%
Power Consumption Using 5 VDC Nominal
  • Sleep: 50 mW
  • Idle: 150 mW
  • Measurement: 1.0 W
  • Inflation: 3.6 W (High)
Communication Protocols TTL (5V)
Safety & Regulatory Standards IEC 60601-1, IEC/EN80601-2-30, EN60601-1-2:2007, ISO 81060-2:2009

Advantage Technologies

Specifications Values
RMT-MAX (Maximum Performance R-wave Motion Tolerance) Yes
TMT-MAX (Maximum Performance Transport Motion Tolerance) Yes
VET-S (Standard Veterinary Monitoring) Yes

Customer Support

Specifications Values
Standard Support Yes
Custom Support 10 hours*
Custom Design Additional Charge

Warranty

Specifications Values
Time 1 year
Cycles 30000

*ADDITIONAL SUPPORT CHARGED AT STANDARD RATES

Looking for a complete list of Parts & Specs for each product? Click here.