Technical Specifications

Physical Dimensions

Specifications Values
Width 1.6 in [40.6 mm]
Length 1.6 in [40.6 mm]
Height 0.71 in [18.00 mm]
Weight 0.25 lbs [113 g]

Measurement

Specifications Values
Technique Oscillometric
NIBP Accuracy Meets ANSI/AAMI SP10, EN 1060-4, ISO 81060-2
Patient Applications Adult, Pediatric, Neonatal
Systolic Range
  • Adult: 40-260 mmHg
  • Pediatric: 40-160 mmHg
  • Neonatal: 40-130 mmHg
Diastolic Range
  • Adult: 20-200 mmHg
  • Pediatric: 20-120 mmHg
  • Neonatal: 20-100 mmHg
Pulse Rate Range 30-220 BPM (Beats per min)
Pulse Rate Accuracy ±2% or ±3 BPM, whichever is greater
Transducer Accuracy ± 3 mmHg over full range in operating conditions
Operating Modes Manual
Operating Conditions 0˚C to 50˚C, 15% to 95%, non-condensing humidity
Storage Conditions -20˚C to +65˚C, up to 90% RH non-condensing

Electrical

Specifications Values
Operating Voltage 5.3-14.5 VDC (High), 1.8-9.0 VDC (Low)
Power Consumption Using 6 VDC Nominal (High) or 3.3 VDC Nominal (Low)
  • Sleep: 2.0 mW (High), 3.0 mW (Low)
  • Idle: 372 mW (High), 430 mW (Low)
  • Measurement: 1.3 W (High), 1.4 W (Low)
  • Inflation: 3.7 W (High), 3.4 W (Low)
Communication Protocols Serial RS232 or TTL (3.3V or 5V)
Safety & Regulatory Standards IEC 60601-1, IEC/EN60601-2-30, AAMI SP10, EN1060-1, EN1060-3, EN1060-4, IEC 80601-2-30, ISO 81060-2

Advantage Technologies

Specifications Values
SMT (Standard Motion Tolerance) Yes
RMT-S (Standard R-wave Motion Tolerance) Yes
TMT-S (Standard Transport Motion Tolerance) Yes
VET (Veterinary Monitoring) Yes
HDM (Hemodialysis Monitoring) Yes

Customer Support

Specifications Values
Standard Support Yes
Custom Support 10 hours*
Custom Design Additional Charge

Warranty

Specifications Values
Time 1 year
Cycles 30000

*ADDITIONAL SUPPORT CHARGED AT STANDARD RATES